Certified Reference Materials for Liquid Density
The following liquid density CRMs, produced on a UKAS accredited hydrostatic weighing system, are normally available ex stock:
|
|
kgm-3 at 20°C
|
Range
|
|
2,2,4-trimethylpentane
|
692
|
15 - 25°C
|
|
dodecane
|
750
|
10 - 50°C
|
|
base lube oil ~8 mPas @ 40°C
|
867
|
15 - 50°C
|
|
base lube oil ~30 mPa s @ 40°C
|
871
|
15 - 50°C
|
|
base lube oil ~110 mPa s @ 40°C
|
881
|
15 - 50°C
|
|
lube oil A90 ~160 mPa s @ 40°C
|
887
|
15 - 50°C
|
|
ethanol in water ~ 41% ABV
|
947
|
20°C†
|
|
ethanol in water ~ 11% ABV
|
984
|
20°C†
|
|
dextrose in water ~ 10% w/w
|
1036
|
20°C†
|
|
dimethylphthalate ~17 mPa s @ 20°C
|
1191
|
15 - 25°C
|
|
sodium bromide in water ~ 30% w/w
|
1206
|
15 - 25°C†
|
|
tetrachloroethylene
|
1623
|
10 - 50°C
|
|
Most standards have an uncertainty of ± 0.01 kgm-3 over the stated temperature range. Ethanol, dextrose, and sodium bromide solutions have an uncertainty of ± 0.03 kgm-3.
Only liquid which has passed through the hydrostatic weighing rig is sold.
Standards are sold in 10 ml sealed glass ampoules.
The displayed density on many density meters is affected by the viscosity of the liquid. If you are measuring viscous liquids, you should consider calibrating your meter with a CRM of similar viscosity.
†We do not recommend the use of aqueous standards unless they are required by the fiscal authorities.
Return to previous page.
|